Supplemental Performance Characteristics; Additional Error - Agilent Technologies 16196A Operation And Service Manual

Table of Contents

Advertisement

Supplemental Performance Characteristics

This section provides useful data on the 16196A/B/C/D. These supplemental performance
characteristics should not be considered specifications.

Additional Error

Additional errors are calculated as follows.
|Z| Measurement
Additional error for Impedance Ze [%] is calculated by substituting the values in the table
below into the following equation.
Ze [%] = ± {A + (Zs/Zx + Yo ×Zx) × 100}
where
A [%]
Yo [S]
Zs [Ω]
Zx [Ω]
Zs
Yo
A
where f is frequency (GHz).
D Measurement
Additional error for Dissipation Factor De is calculated by using the additional error for
Impedance Ze [%] as follows.
If Dx ≤ 0.1:
If 0.1 < Dx ≤ 0.5:
where Dx is the measured value of D. It is necessary for Ze to be below 10 %.
NOTE
D is not expressed as a percentage but as an absolute value.
Chapter 5
Specifications and Supplemental Performance Characteristics
Supplemental Performance Characteristics
Test Fixture's Proportional Error [%]
Test Fixture's Open Repeatability [S]
Test Fixture's Short Repeatability [Ω]
Measured Impedance Value of DUT [Ω]
-3
(30 + 125 × f ) × 10
[Ω]
-6
(5 + 40 × f ) × 10
[S]
2
1 × f
[%]
De = Ze / 100
De = (Ze / 100) × (1 + Dx)
61