Configuration

In Chemstation, left-click the Chip Cube icon or select More Chip Cube from the
Instrument menu. Click Configuration to access the configuration dialog box.
Figure 21

Configure Loading Pump and Analytical Pump

Before working with the system, it is highly recommended to configure both
loading pump and analytical pump for the Chip Cube. When configured, the
pumps react to the status of the Chip Cube and turn the flow off when the chip
is not in Operate (spraying) position. This ensures that no solvent is spilled into
the Chip Cube while the chip tip is retracted or when the chip is unloaded and
the valve stator ports are open. In addition, it protects the enrichment column
from damage by uncontrolled pressure release.