You can access additional functions by right-clicking anywhere in the CM-30S
control panel.
Additional functions in the CM-30S control panel:
Control panel function
Launch instrument console
Reset CM
Help

Installation recommendations for fittings

Warning:
fittings contaminated with biologically hazardous or toxic
materials, always wear clean, chemical-resistant, powder-free
gloves when reinstalling the fittings.
The system uses gold-plated compression screws and two-piece ferrules. See
the diagram below for assembly orientation.
Ferrule with
locking ring
Recommendations:
To prevent bandspreading, ensure the tubing bottoms in its fitting hole
before tightening the compression screw.
For easier accessibility, use long compression screws to attach tubes to
the injector and vent valve.
Whenever you loosen fittings during maintenance, examine them for
cracks, stripped threads, and deformations.
Do not reuse stainless steel fittings more than six times.
To avoid the harmful effects of personal contact with
Compression screw

Installation recommendations for fittings

Description
Launches the instrument console.
Resets the CM-30S after an error
condition.
Displays the online Help for the
instrument console software.