Resolving leak sensor errors

After approximately 1.5 mL of liquid accumulates in the leak-sensor reservoir,
an alarm sounds, indicating that the leak sensor detected a leak.
Warning:
biologically hazardous or toxic materials, always wear clean,
chemical-resistant, powder-free gloves when performing this
procedure.
Notice:
• do not allow buffered solvents to accumulate and dry on it;
• do not submerge it in a cleaning bath.
Small column leaks can be undetectable because they can evaporate
Note:
before reaching the leak sensor reservoir.
Required materials
Cotton swabs
Gloves: clean, powder-free, chemical-resistant
Nonabrasive, lint-free wipes
To avoid electric shock, do not remove the instrument's panels.
Risk of tipping. To avoid injury, ensure the side-mount shaft
for safety advisory information.
To avoid personal contamination from contact with
To avoid scratching or damaging the leak sensor,

Resolving leak sensor errors

October 7, 2014, 715004178 Rev. B