Performance Specifications - Waters ACQUITY CM-30S Overview And Maintenance Manual

30-cm single zone column manager
Table of Contents

Advertisement

Electrical specifications: (Continued)
Attribute
Voltage range
Frequency
Maximum power draw
a. Protection Class I
electrical shock. Class I identifies a single level of insulation between live parts (wires) and
exposed conductive parts (metal panels), in which the exposed conductive parts are connected
to a grounding system. In turn, this grounding system is connected to the third pin (ground
pin) on the electrical power cord plug.
b. Overvoltage Category II
a local level such as an electrical wall outlet.
c. Pollution Degree 2
reduction of dielectric strength or surface resistivity. Degree 2 refers only to normally
nonconductive pollution. Occasionally, however, expect a temporary conductivity caused by
condensation.
d. Moisture Protection – Normal (IPX0) – IPX0 means that no Ingress Protection against any
type of dripping or sprayed water exists. The "X" is a placeholder that identifies protection
against dust, if applicable.

Performance specifications

The following table lists the performance specifications for the CM-30S.

Performance specifications:

Item
Analytical columns
Guard columns
Specification
100 to 240 VAC
50−60 Hz
600 VA
The insulating scheme used in the instrument to protect from
Pertains to instruments that receive their electrical power from
A measure of pollution on electrical circuits that can produce a
Performance specifications
Specification
ID: 2.1 to 8.0 mm
Length: 30.0 to 300.0 mm
ID: 2.1 to 8.0 mm
Length: 50.0 mm maximum
October 7, 2014, 715004178 Rev. B