RS-232 Parameters
Note:
RS-232 parameters are not user-configurable.
Table 2-5 RS-232 Parameters
Parameter
Baud rate
1200
Number of bits
8
Number of stop bits
1
Parity
none
2.6 Changing the Hardware Configuration
This section covers:
• Removing the autosampler cover
• Installing the heater/cooler
• Installing the auxiliary sample loop

2.6.1 Removing/Replacing the 717plus Cover

Caution:
To avoid the possibility of electric shock, disconnect the autosampler power cord
before continuing this procedure.
Removing the Cover
To remove the autosampler cover
1. Turn off the autosampler and disconnect the power cord.
2. Remove the syringe cover, and unscrew the cover retaining screw located to the
right of the magnetic latch.
3. Slide back the cover until it clears the posts on the rear panel.
4. Ensure that the red and blue fluid lines are free of the side cover slots.
5. Lift the cover up and back.