Chapter 5 Maintenance - Waters 717plus Operator's Manual

Hide thumbs Also See for 717plus:
Table of Contents

Advertisement

5
Maintenance
This chapter covers 717plus maintenance. Procedures are provided for defrosting the
heater/cooler and for replacing the syringe, seal pak, and in-line filters.
5.1 Defrosting the Heater/Cooler
When to Defrost
If the Heater/Cooler fails to reach the lowest setpoint at recommended ambient conditions
(16 to 27° C and 80 percent relative humidity), defrost the heater/cooler compartment.
Note:
Maintain proper ventilation space (see
Defrosting the Heater/Cooler
To defrost:
1. Press the keys outlined in the figure below
2. Open the sample compartment door. The defrost cycle takes 60 minutes.
Note:
If you want to stop the defrost cycle before the programmed time expires, press
Stop
Defrost.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Section 2.1, Site
Requirements).
(Figure
5-1).
Defrosting the Heater/Cooler