2.3 Changing Voltage Configuration
The Waters 717plus autosampler is shipped in a 120 V/60 Hz configuration. If you require
a different voltage setting, change the configuration.
Tools required:
• One small-blade screwdriver
• One Phillips-head screwdriver
• One pair of needlenose pliers
2
Caution:
Disconnect the autosampler power cord before continuing this procedure.
For continued protection, use only fuses of the approved type and rating, as
indicated by the label on the rear panel near the power cord connector
Selecting Voltage
To select the voltage configuration:
1. Disconnect the power cord from the back of the Waters 717plus Autosampler.
2. Insert a small-blade screwdriver into the notch just to the right of the power cord
connector and pry off the fuse block cover
Figure 2-2 Fuse Block and Voltage Selector Card
3. Grasp the voltage selector card and pull it straight out of the housing.
26
Installation
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
(Figure
Notch
Voltage Selector
Card
Voltage indicated
by white pin
Fuse Block
Fuse Block Cover
2-2).