To stop a Stat run:
1. Press
Stop
Run. A prompt appears:
Do you really want to stop the Auto Run?
2. Press
Yes
to stop the run and return to the Main Menu.

4.8 Sampling Considerations

The sample and chromatography method you use may require changing the configuration
of the autosampler. This section covers:
• Setting the temperature control for the sample compartment
• Making microbore (less than 5 µL) or semiprep (greater than 250 µL) injections
• Injecting viscous samples
• Purging during an auto run
• Using buffered mobile phases

4.8.1 Temperature Control

The heater/cooler module provides sample compartment temperature control from
4 to 40° C.
Environmental Restrictions
Ambient temperature must be 16 to 27° C at less than 80 percent relative humidity. At
higher ambient temperatures, cooling capacity may be degraded.
Note:
Ensure that ventilation space is maintained (see
To control sample compartment temperature:
1. When changing set point from below ambient to above ambient temperature, defrost
the heater/cooler to prevent moisture build up inside the sample compartment (see
Section 5.1, Defrosting the
2. Set the sample chamber temperature on Configuration Page 2 by pressing the
keys indicated in
Figure
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Section 2.1, Site
Heater/Cooler).
4-25.
Sampling Considerations
4
Requirements).
116