Microbore/Semiprep Injections - Waters 717plus Operator's Manual

Hide thumbs Also See for 717plus:
Table of Contents

Advertisement

Figure 4-25 Setting Sample Chamber Temperature
3. Close the sample compartment door.
4
Note:
The heater/cooler shuts off if the sample compartment door is left open more
than 5 minutes
4. Wait for the sample compartment to reach the set point. The heater/cooler may
require up to 90 minutes to reach 4° C from an ambient temperature of 27° C.

4.8.2 Microbore/Semiprep Injections

The Waters 717plus a utosampler may be physically reconfigured to facilitate microbore
injections (< 5 µL) and semiprep injections (> 150 µL).
117
Operating the Autosampler
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Menu: Main
Config
Page
Configuration
Page 1
NextCfg
Page
IDLE
Edit: Configuration
Page 2
Hold Contact Sense?
+
-
Hold Input?
On
Off
Inject Start Pulse (sec):
0.5
End of Run Pulse (sec):
0.5
Temp Setpoint(¡C):
8
Temp Control?
On
Off
On
PrevCfg
Main
NextCfg
Page
Page
Page