Cleaning/disinfecting the SiroCam AF / AF+ intraoral camera

The shape of the SiroCam AF / AF+ intraoral camera complies with
hygienic requirements and therefore has no areas that are difficult to
reach. It can be wipe disinfected.
NOTE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning, and disinfecting agents that have been
approved by Dentsply Sirona, see "Care, cleaning, and disinfecting
agents"  [→ 203].
NOTE
The lens window is sensitive to scratches.
Deep scratches in the lens window impair image quality.
➢ Protect the lens window against scratching. Disinfect it with a soft,
lint-free cloth.
5.3.4.5
Check the flow rate on the Sprayvit M multifunctional syringe
If the flow rate is less than 135 ml/min with the button fully actuated,
there is the risk of excessively hot water being emitted.
CAUTION
If the flow rate is insufficient, hot water may be emitted by the Sprayvit
M.
The patient could thus be scalded.
➢ Check the water flow rate prior to use.
➢ Clean the nozzle according to the Sprayvit M Operating
Instructions.
IMPORTANT
Air outlet during hose change
If the Sprayvit M is removed from the instrument hose while the
treatment center is in use, air will escape from the hose coupling. You
should therefore switch off the treatment center before changing a
hose.
Perform the following measurement to exclude risk to the patient.
➢ Fill a measuring cup up to the 100 ml mark with the water button
fully actuated while measuring the required filling time.
Ä The filling time must not exceed 45 seconds.
If the specified water quantity is not reached within the filling time of 45
seconds, clean the nozzle or have the treatment center checked by a
service technician.
5 Care, cleaning and maintenance by the practice team
5.3 Instruments and instrument hoses
227