4.1.2.2

Main switch

Switching the treatment center on
The main switch switches the treatment center from Standby mode to
operational readiness.
Following switch-on, the operating system is booted and an automatic
self-test is performed.
ü The power switch is turned on.
➢ Press the main switch on the dentist element.
Ä The LED of the main switch lights up on the dentist element.
Ä The treatment center powers up and establishes operational
readiness.
If the next maintenance call is due in less than 42 days or the
maintenance deadline has already been exceeded, a message appears
on the touchscreen. For more information, please refer to "Inspection
and maintenance"  [→ 265].
Switching the treatment center to the Standby mode
On completing your work, you should switch the treatment center off
with the main switch on the dentist element both for safety reasons and
to reduce its power consumption. Pressing the main switch turns the air
and water supply as well as all electronic components off. Only the
Standby circuit is still supplied with voltage. If no other devices are
operating using the external device connection and the treatment center
is not equipped with an internal PC, the power consumption in Standby
mode is approximately 3 W.
➢ Press and hold the main switch on the dentist element until an
acoustic signal sounds. Then release the key.
Ä The treatment center then shuts down and switches itself to the
Standby mode.
Ä The LED of the main switch goes out on the dentist element.
4 Operation
4.1 Starting up the treatment center
53