Setting The Apexlocator; Switch Instrument Light On/Off; Setting The Foot Control As A Direct Starter Or Speed Foot Control - Dentsply Sirona Sinius Operating Instructions Manual

Hide thumbs Also See for Sinius:
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions Sinius
S
63  2 2  7 26 D3561
D3561.201.01.16.02    08.2018
Via touchscreen
The preselected coolant in the EasyMode Start program operating
mode can only be turned on/off with the foot control buttons, see below.
ü An instrument is taken from the holder and the Instrument program
is displayed on the touchscreen.
➢ Touch the key of the preselected coolant ( Spray , Air or NaCl ).
Ä If the key of the preselected coolant is highlighted orange, it will
be switched on together with the instrument when the foot pedal
is activated. If the key is highlighted gray, the coolant is
switched off.
Via foot control
ü An instrument is removed from its holder.
➢ Press the left button of the foot control.
Ä If the key of the preselected coolant ( Spray , Air or NaCl ) is
highlighted orange on the touchscreen, it will be switched on
together with the instrument when the foot pedal is activated.
4.5.8.4

Setting the ApexLocator

If your treatment center is equipped with the ApexLocator option, you
can adjust this in the Motor sub-screen.For more information, refer to
the chapter "ApexLocator"  [→ 103], section "Endodontic treatments with
motor". For more information, refer to the chapter "ApexLocator" 
[→ 103], section "Implantology/endodontic treatments with motor" 
[→ 117].
The ApexLocator can only be used in the Standard Start program
operating mode.
4.5.8.5

Switch instrument light on/off

ü The sub-screen of the removed turbine of a motor or scaler is
displayed on the touchscreen.
➢ Switch the instrument light on or off with the Instrument light key.
Ä If the key is highlighted orange, the instrument light can be
activated using the foot pedal.
For high-speed handpieces, the brightness and operating voltage of the
instrument light can be set, see "Setting the handpiece light"  [→ 120].
The instrument light of the Sprayvit M multifunctional syringe is set
separately, see "Switching the instrument light on/off and setting the
water temperature"  [→ 123].
4.5.8.6

Setting the foot control as a direct starter or speed foot control

The foot control can be set as a direct starter or as a speed foot control
in a motor or SiroSonic TL scaler sub-screen:
● Direct starter
When the foot control is actuated, the instrument is switched on with
the set speed and intensity.
● Control foot control
4 Operation