Implantology/Endodontics Treatments With Motor - Dentsply Sirona Sinius Operating Instructions Manual

Hide thumbs Also See for Sinius:
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions Sinius
63  2 2  7 26 D3561
D3561.201.01.16.02    08.2018
4.5.10.4

Implantology/endodontics treatments with motor

4.5.10.4.1
Electronic torque limitation
If your treatment center is equipped with the implantology/endodontics
software option, please observe the instructions provided in the chapter
"Implantology and endodontic treatments"  [→ 129]. If your treatment
center is not equipped with this option, please note that you do not have
the option of electronic torque limitation.
CAUTION
No electronic torque limitation is available in the Motor program.
Root canal files can easily break during operation without torque
limitation.
➢ Never perform endodontic treatments without torque limitation.
Use an endodontic handpiece with mechanical torque limitation,
e.g. SiroNiTi from Dentsply Sirona.
4.5.10.4.2
Endodontic treatments with the ApexLocator in the motor screen
If your treatment center is equipped with the ApexLocator option but
does not have the treatment function, the Apex function in the Motor
program can be used for endodontic treatments.
Please note that the ApexLocator can be accessed only in the Motor
program in the Standard Start program operating mode.
IMPORTANT
SiroNiTi Apex
Only Dentsply Sirona SiroNiTi Apex must be used to perform apex
measurement using a torque-limiting handpiece! The electrical
conductivity can be ensured only using this handpiece.
The SiroNiTi handpiece can also be used to perform endodontic
treatments without the ApexLocator.
ü The treatment center is prepared for apex measurements using the
instrument, see section "Preparing to use the ApexLocator"  [→ 104].
ü The Motor dialog is displayed on the touchscreen.
1. Set the appropriate motor speed according to the handpiece and
the root canal file used, see "Setting the speed"  [→ 115].
2. You can set the motor so that it stops automatically when the
physiological apex is reached. If you want to use the automatic
motor stop, this can be preset in the Motor sub-dialog. To do this
touch the Sub-dialog key.
Ä The Motor sub-dialog is displayed.
3. Touch the Apex stop key.