To Change The Maintenance Expel Time; Disposal; Cleaning And Disinfecting; Replacing Fuses - Dentsply Sirona MIDWEST AUTOMATE Instructions For Use Manual

Automated handpiece maintenance system
Table of Contents

Advertisement

Available languages

Available languages

5 Maintenance

To change the maintenance expel time

Turn power off.
Press and hold
.
Turn power on.
The LED
(ready to operate) flashes.
Use
to change settings.
Maintenance time is set as follows:
15 sec
LED 1 illuminates
30 sec
LED 2 illuminates
45 sec
LED 3 illuminates
60 sec
LED's 1, 2 and 3 illuminates together
When time setting is chosen, wait 5 seconds for change to take effect.
After which the unit assumes normal operation (LED
NOTE:
Factory default setting is 60 seconds.

Disposal

See "Disposal of Consumables" (page 3).

Cleaning and disinfecting

Clean and disinfect powder coated paint surfaces with common disinfectants (Dentsply Sirona Volowipes, SciCan
OPTIM wipes, Metrex CaviWipes).
Clean operating keypad with soft cloth dampened with water.

Replacing fuses

Disconnect power.
Pull out fuse drawer.
(Push both flaps of the fuse drawer inward.)
Replace fuses.
Insert fuse drawer.
®
Use only Midwest
supplied fuses.

Replacing front door seal

Remove damaged door seal (at least one complete side).
Clean the bonding surface with rubbing alcohol.
Cut door seal to length.
Place new adhesive door seal flush with the inner edge of door.

Replacing rubber foot

Remove rubber foot.
Using the same screw and washer attach new rubber foot.

Replacing air tube

Remove tube from wall air supply.
Press and hold release ring on top of the compressed air coupling (page 5) and pull air tube out.
Slide a new tube on to the compressed air coupling.
Attach opposite end of tube to wall air supply.
illuminates).
M
IDWEST
10
Automated Handpiece Maintenance System
®
A
®
UTOMATE
EN

Hide quick links:

Advertisement

Table of Contents
loading