Calibrating the inEos X5 camera

IMPORTANT
Calibrating the system
The calibration of the inEos X5 camera is only required if you use
ATLANTIS-FLO-S scanbodies in order to design single-piece directly
screwed suprastructures and to have them manufactured at
INFINIDENT SOLUTIONS or in order to transmit the scanned models
to Dentsply-Sirona-ATLANTIS central manufacturing for design and
manufacturing purposes.
ü The inEos X5 is switched on.
ü The optionally available calibration set inEos X5 (REF 6483759) is
available.
ü You have restarted the software.
1. Click the "Devices" button in the system menu.
2. Click on "inEos X5" .
Ä A selection menu opens.
3. Click on the "Calibrate Camera" button.
Ä The calibration dialog opens.
4. Once prompted, position the calibration set on the rotary plate as
shown in the software.
5. Start the calibration process.
Ä The unit is calibrated.
Ä Depending on the ambient temperature, the camera optical
system may need some time to warm up. This is required in
order to achieve maximum precision.
6. Once the calibration is complete, the calibration protocol can be
saved in PDF format.
7. Restart the software.
6 Initial startup
6.5 Calibrating the inEos X5 camera