8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
60

Operating Instructions (valid for Canada) CEREC AF, CEREC AF Connect

Disinfecting the camera body and mirror sleeve

8.1.3.7
NOTICE
Do not spray the CEREC Omnicam or immerse it in cleaning agents or
disinfectants!
➢ Use a new cotton gauze moistened with one of the cleaning products
listed in the „Cleaning and disinfection agents [ → 56]" section. Wipe
the camera body first and then the mirror sleeve. This step disinfects
the camera body and the mirror sleeve. Then dry the mirror sleeve
with a clean lint-free cloth.
8.1.3.8

High-level disinfection of the mirror sleeve (with HLD Set)

NOTICE
Dry heat sterilization and high-level disinfection must not be combined.
The complete process for high-level disinfection is as follows:
With HLD Set
A HLD set to support the HLD process should be ordered from Dentsply
Sirona with REF 66 05 120.
A
HLD container
B
Lid
1. Pre-cleaning (see „Pre-cleaning of the mirror sleeve [ → 59]").
2. Put on Personal Protective Equipment at the latest when handling the
high-level disinfectant.
3. Remove the mirror sleeve from the CEREC Omnicam (see
„Removing the mirror sleeve [ → 58]").
C
Metal holder
D
Plug
D3652.201.01.14.02 10.2017
Dentsply Sirona
66 37 487 D3652