2 Safety instructions

2.3 On-site installation

16
On-site installation
2.3
The on-site installation must have been performed according to our
requirements. The details are described in the document "Installation
Requirements".
2.4

Installation of the treatment center

Installation must be carried out by authorized personnel according to
the installation instructions.

Media quality

2.5
The air and water supplies must meet the requirements specified in the
installation requirements.
To ensure compliance with the medical and national legal requirements
for water from treatment centers, Dental Sirona recommends equipping
the treatment center with a disinfection system. If you decide to operate
the treatment center without a disinfection system, you must make
alternative arrangements to ensure good water quality.
As the owner of the treatment center, you are generally responsible for
the water quality.
The bacterial count must comply with the national regulations for
drinking water and must not exceed 500 CFU/ml under any
circumstances (CFU: colony forming unit).
If the bacterial count is too high, the building water system must be
checked and the cause of contamination eliminated. Alternatively, a
self-sufficient water supply can also be installed. In addition, if the
treatment center is equipped with a disinfection system, the empty
disinfectant tank can be used as a water container.
Before installation of the treatment center, an acceptable
microbiological water quality for the domestic water supply must be
ensured and documented in the form of the bacterial count. Sampling
and the bacterial count must be performed by a competent laboratory.
Test the water quality from the treatment center at regular intervals and
after it has not been used for >1 week, see "Microbiological water test" 
[→ 203]. Please contact your specialized dealer or your dental
association for the respective national requirements and measures.
Where necessary, you must make alternative arrangements to ensure
good water quality if you decide to operate the treatment center without
a disinfection system.
WARNING
Highly immunosuppressed patients or patients with specific
pulmonary diseases should not come in contact with the water of the
treatment unit . The use of sterile solutions is recommended.