Operating Instructions Midwest Stylus
Plus Air-Driven-High-Speed Handpiece
7 Conditioning
Please observe the manufacturer's instructions for using instrument
disinfectants.
7.2

Automated cleaning and disinfecting ...

NOTICE! Do not carry out automated cleaning of the quick couplings.
Quick couplings excluded
7.2.1
... with cleaning and disinfection equipment
The handpiece can also be cleaned and disinfected in suitable cleaning
Sirona _ -Handpiece
and disinfection equipment.
The cleaning and disinfection equipment used must be approved by its
Cleaning and disinfection equipment
manufacturer for the cleaning and disinfection of dental instruments and
comply with EN ISO 15883-1 (e.g. 95°C (203°F) and 10 min. holding
time).
For further details, refer to the operating instructions supplied with the
unit.
✔ The handpiece is conditioned with a cleaning and disinfection
Checking _ Handpiece
device.
39
6607290 D3673.201.01.01.02