5 Care, cleaning and maintenance by the practice team

5.2 Surfaces

206
Surfaces
5.2
5.2.1

Clean/disinfect surfaces

The surfaces can be disinfected by spraying and wiping with surface
disinfectants.
NOTE
Drugs have a chemical reaction with the surface of the unit.
Due to their high concentrations and the substances they contain,
many drugs can dissolve, etch, bleach, or stain surfaces.
➢ Wipe any drug residues off the unit immediately with a moist,
white cloth!
NOTE
Liquids can enter the unit during cleaning or disinfection.
Electrical components of the treatment center can be destroyed by
liquids.
➢ Do not spray any liquids into the unit.
➢ To clean near openings, first spray the liquid onto a cleaning
cloth. Then wipe over the unit with the cleaning cloth.
NOTE
Disinfectants can dissolve dyes in cleaning clothes.
The outer surface of the unit may then be discolored by the dye.
➢ Do not clean or disinfect the unit with colored cleaning cloths.
➢ Remove any dirt and disinfectant residues regularly using a mild
commercial cleaning agent.
Dentsply Sirona
Operating Instructions Sinius
63  2 2  7 26 D3561
D3561.201.01.16.02    08.2018