NOTE
Approved care, cleaning, and disinfecting agents
Use only the agent for disinfecting the water paths that has been
approved by Dentsply Sirona for the disinfection system; see "Care,
cleaning, and disinfecting agents"  [→ 203].
5.5.5

Change the water and air filters

If you notice any changes in media flow rates, check the water and air
filters for permeability. The filters must be changed, if necessary.
To reorder the filters, see "Spare parts and consumables"  [→ 270].
1. Switch the treatment center off at the main switch.
Ä The water and air supply are switched off.
2. Open the maintenance flap A.
3. Any remaining water flows out of the water filter when it is opened.
You should therefore place an absorbent cloth underneath the filter.
Then unscrew and remove the screw cap of the water filter B and/or
the air filter C.
4. Check the filters and replace them if necessary.
5. Put the sealD back on the filter E. Then insert both into the screw
cap B, C as shown.
Ä The filter E clicks into the screw cap B, C.
6. Screw the screw-on cap(s) B, C back onto the water unit.
5 Care, cleaning and maintenance by the practice team
5.5 Components of the water unit
241