Connecting The Mini L.e.d; Functional Description - Dentsply Sirona Sinius Operating Instructions Manual

Hide thumbs Also See for Sinius:
Table of Contents

Advertisement

4 Operation
4.6 Assistant element
E
D
C
A
156
4.6.6.3

Connecting the Mini L.E.D.

The Mini L.E.D. is connected to the assistant element in holder 1.
B
1. Connect the supply cable A to the treatment center.
2. Screw hose coupling B onto the Mini L.E.D. C.
3. Screw the sterilized light guide D onto the Mini L.E.D. C. Make sure
to insert the fiber optic correctly.
Ä The fiber optic clicks into place audibly.
4. Slip the glare shield E onto the light guide D.
Ä The glare shield protects your eyes against reflecting curing
4.6.6.4

Functional description

Operational Elements
A
A
Glare shield
B
Fiber
C
On/off button
D
Status control lamp
E
Quick curing mode light
Status control lamp
The status control lamp D signals the following states:
Status control lamp
OFF
Green
Red flashing
The Mini L.E.D. has three operating modes, which you can select with
the Mode H key:
light.
B
C
D
Mode
Instrument in holder
Normal mode
Overheat protection
Dentsply Sirona
Operating Instructions Sinius
E
G
I
F
H
F
Pulse curing mode light
G
"Soft start" mode lamp
H
Mode button
I
Handpiece
J
Supply cable
63  2 2  7 26 D3561
D3561.201.01.16.02    08.2018
J

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sinius tsSinius cs

Table of Contents