Machine disinfection
Thermal disinfection at up to 93°C (200°F) according to ISO 15883-1 is
possible for all marked components. Use an alkaline cleaning agent. A
laboratory dishwasher (e.g. Miele G7781) can be used as disinfection
device. See the operating instructions for the respective device for the
setting of the disinfecting equipment and the kind and quantity of
disinfectant.
Manual drying
With the wipe disinfection method, drying is not necessary because
excess disinfectant evaporates. Excess water from the cleaning process
can be removed with a soft cloth.
Sterilization
Sterilization may be conducted for components that are marked
accordingly. Dentsply Sirona recommends sterilization in a steam
sterilizer (autoclave) according to ISO 13060, Class B (e.g., DAC
PREMIUM / DAC PROFESSIONAL).
The sterilization must be carried out with a multiple fractionated suction
system (Class B sterilizer). The process parameters can be found on
the embossed marking of the respective component and this operating
manual. In addition, please follow the operating instructions for the
sterilizer.
5.1.5

Inspection, maintenance and testing

Unless otherwise specified in this operating manual, test all components
for proper functioning on a regular basis and carry out a visual
inspection for damage and wear. Exchange damaged components if
necessary.
5 Care, cleaning and maintenance by the practice team
5.1 Basics
205