Malfunctions; Error Messages - Dentsply Sirona Sinius Operating Instructions Manual

Hide thumbs Also See for Sinius:
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions Sinius
Code
Anomaly
10
Flow rate of the Sprayvit
M on dentist element too
low
11
Flow rate of Sprayvit M on
assistant element too low
12
Tumbler temperature fuse
defective
13
Battery of wireless foot
control empty
14
Consumption of the agent
for disinfecting the water
paths too low
15
Water feed too low
17
Emergency pump opera-
tion
18
Chair movement outside
of the permissible move-
ment limits or chair drive
position not plausible
19
Injection valve for the
agent for disinfecting the
water paths defective
63  2 2  7 26 D3561
D3561.201.01.16.02    08.2018
7

Malfunctions

Error messages

7.1
Error states of the treatment center that cannot be immediately
recognized by the user but must nevertheless be corrected are
displayed on the touchscreen. If an error state exists, this is indicated by
a warning triangle A in the status line of the touchscreen.
➢ Touch the warning triangle A in the status bar of the touchscreen.
Ä The error code(s) is(/are) displayed.
The error codes have the following meaning:
Description
The water flow rate is below the
limit. The heater of the Sprayvit M is
no longer heated to prevent scald-
ing by overheated water.
The temperature fuse was blown by
overheating. The water in the tum-
bler filling unit is no longer heated.
Operation of the treatment center no
longer possible due to fully depleted
battery.
The disinfectant tank was not emp-
tied since its last filling even though
many disinfection cycles were per-
formed.
The maximum filling time for the
mixing tank is exceeded.
Malfunctioning of pump sensors in
the water unit.
If the chair is located outside of the
permissible movement limits or the
position is not plausible, the chair
stops moving.
No agent for disinfecting the water
paths will be added if the injection
valve is defective.
7 Malfunctions
7.1 Error messages
Measure
Clean the outlet nozzle of the
Sprayvit M, see the "Sprayvit M op-
erating instructions and perform the
flow test, see "Checking the flow
rate on the Sprayvit L multifunc-
tional syringe"  [→ 227].
Inform your service technician.
See "Changing the battery of the
wireless foot control"  [→ 262].
Inform your service technician
Change the water filter, see "Chang-
ing the water and air filters"  [→ 241].
Check the water pressure of the
drinking water supply.
Inform your service technician.
Inform your service technician.
Inform your service technician.
267