5 Care, cleaning and maintenance by the practice team

5.6 Sanitation

250
Sanitation
5.6
Sanitation effectively combats the reproduction of microorganisms in the
water paths.
If the disinfection system is set for operation with public drinking water
supply, the sanitization must be guided using the sanitization screen,
see "Interactively guided sanitation of the treatment center"  [→ 250].
If the equipment is being operated with a self-sufficient water supply,
sanitation can only be carried out manually, see "Sanitizing the
treatment center manually"  [→ 257]. The Sanitation program is not
available in this operating mode.
For more information, please refer to the section entitled "Self-sufficient
water supply"  [→ 161].
5.6.1

Interactively guided sanitation of the treatment center

Sanitization, that is disinfecting the treatment water paths, can be
performed by using the disinfection system. During sanitization, the
treatment water is first emptied according to a pre-defined process, then
the disinfectant for the water paths is filled undiluted into the treatment
water paths and flushed out again at the end. Sanitization includes
several phases and takes at least 24 hours but should not exceed 3
days.
When operating with public drinking water supply (self-sufficient water
supply switched off), interactive sanitization should be carried out:
● Regularly every 4 weeks
The display Days until next sanitization appears in the status bar of
the touchscreen. It first appears three days before the date of
sanitization (3d = 3 days until sanitization).
● After longer periods of disuse (> one week)
● If the bacterial count exceeds 100 bacteria per milliliter, see
"Microbiological water test"  [→ 203].
● After changing from stand-alone water supply to public drinking
water supply and the agent for disinfecting the water paths
NOTE
Additional devices connected to the external device connection must
not be sanitized with the treatment center.
The additional devices could be damaged. Residues of the agent for
disinfecting the water paths can remain in the additional devices.
➢ Disconnect any additional devices from the treatment center prior
to sanitization.
Preparation
The following preparations should be made prior to beginning
sanitation.
1. If your treatment center is equipped with a cuspidor, activate the
cuspidor flushing for at least one minute. This flushes the water
paths.
Dentsply Sirona
Operating Instructions Sinius
63  2 2  7 26 D3561
D3561.201.01.16.02    08.2018