4.4.3

Triggering an immediate movement stop

You can stop the movement of the chair to a programmed position as
follows:
➢ Press the main switch or the Shock positioning fixed key on the
dentist element.
➢ Press a button relating to the patient chair on the touchscreen or a
button to tilt the motor-driven headrest.
➢ Press one of the patient chair keys on the control panel of the
assistant element.
➢ With all instruments are in place press the pedal or the left or right
button of the foot control.
➢ With an instrument removed, press the pedal of the foot control.
➢ With the cursor control switched off, actuate the 4-way foot control
plate on the foot control in any direction.
➢ With the cursor control switched on, move the cursor to any of the
keys relating to the patient chair.
➢ Move the 4-way foot switch in any direction.
➢ Actuate one of the 4-way switches on the motor-driven headrest in
any direction. Exception: The upper 4-way switch up/down, see
note below.
Ä All movements of the treatment center are stopped immediately.
IMPORTANT
The function keys and switches for adjusting the headrest to the
height of the patient can be used during the program run. This does
not trigger a movement stop.