4.6.6.3

Connecting the Mini L.E.D.

The Mini L.E.D. is connected to the assistant element in holder 2.
B
1. Connect the supply cable A to the treatment center.
2. Screw hose coupling B onto the Mini L.E.D. C.
3. Screw the sterilized light guide D onto the Mini L.E.D. C. Make sure
to insert the fiber optic correctly.
Ä The fiber optic clicks into place audibly.
4. Slip the glare shield E onto the light guide D.
Ä The glare shield protects your eyes against reflecting curing
4.6.6.4

Functional description

Operational Elements
A
A
Glare shield
B
Fiber
C
On/off button
D
Status control lamp
E
Quick curing mode light
Status control lamp
The status control lamp D signals the following states:
Status control lamp
OFF
Green
Red flashing
The Mini L.E.D. has three operating modes, which you can select with
the Mode H key:
Quick curing mode
In the quick curing mode, the Mini L.E.D. operates at full power for 10
seconds.
In this mode, the light intensity output is as follows:
● 1250 mW/cm
● 2000 mW/cm
light.
B
C
D
Mode
Instrument in holder
Normal mode
Overheat protection
²
(± 10%) with the standard light guide, dia. 7.5 mm
²
(± 10%) with the booster light guide, dia. 5.5 mm
Dentsply Sirona
Operating Instructions Intego
E
G
I
F
H
F
Pulse curing mode light
G
"Soft start" mode lamp
H
Mode button
I
Handpiece
J
Supply cable
64  5 7  4 98 D3543
D3543.201.01.08.02    02.2020
J