5 Installation and startup
5.4 Commissioning
26
5.4.6

Installation

5.4.6.1

Connecting to the PC via LAN

An Ethernet port is located on the rear of the unit, which can be used to
connect the PC to the production unit. Use a network cable to do this
(LAN connection).
Using a network cable
Connect the PC to the LAN connection of the unit.
5.4.6.2

Connecting the unit to the power supply

NOTE
Grounded power outlet
The unit must be connected to a grounded power outlet.
➢ Connect the unit to the power supply using the supplied power
cable.
5.4.6.3

Installing the unit

The following steps must be performed before putting the unit into
operation.
5.4.6.3.1
Process steps that have to be taken on the unit.
ü The unit is connected to a power supply source and the main switch
on the rear of the unit is switched on.
1. Switch the unit on by pressing the On/Off button at the front.
Ä A start screen will shortly appear on the touch display.
2. Select the set-up language.
Ä The unit looks for a hub or a virtual software hub.
3. If using a virtual software hub solution, make sure that the PC on
which the software hub is executed is switched on.
Ä The touch display displays information if the hub connection
has been established.
4. If no hub or software hub is available, you can change the IP
settings of the CEREC Primemill without continuing the hub or
performing a manual hub search.
5. Check the hub presets and change them if required.
6. Check on the touch display whether or not an suction unit is
connected.
Dentsply Sirona
Operating Instructions CEREC Primemill
67 19 681 D3439
D3439.201.01.01.02    09.2019