Chapter 11 FiberComplete Modules

Results screen

d
e
See
"Configuring the units" on page
Results screen
Once the tests are completed, the results screen displays on both units.

Cable view

If, in the Setup page, the parameter Table View is set to Cable, the following result table
is displayed:
Figure 130
Result Cable View
The Cable View allows to display results of multiple fibers:
210
Once all tests are performed, results are displayed on both units
If OTDR parameter is set to Auto or Manual in the Setup menu, the OTDR
acquisition starts.
If Fault Finder parameter is set to Yes and an IL or ORL value reaches one
of the user defined thresholds, the Fault Finder starts.
the average loss at each wavelength
the ORL A and ORL B at each wavelength
the distance is displayed on top of the screen