Figure 125
FiberComplete Setup
See
"Configuring the
File parameters" on
page 44
Acquisition parameters
•
Laser
NOTE
It is possible to select individual wavelength per unit. For example, one can
make 1490/1550 nm test from one end, and 1310 nm from the other end.
The following acquisition parameters are only taken into account if the
test is performed from this unit. If not, those parameters are not taken
into account.
•
Loss measurement
Unidir.: loss is measured in one direction only
Bidir.: loss is measured in both direction
None: loss is not measured.
•
ORL measurement
Unidir.: ORL is measured in one direction only
Bidir.: ORL is measured in both direction
select the desired wavelength(s).
All: acquisition is performed for all wavelengths available in the
instruments.
select the loss measurement mode.
select the ORL measurement mode.
User Manual
Chapter 11 FiberComplete Modules
77000010208 Rev. 012
Configuring the units
See
"Configuring the
Link parameters" on
page 39
201