Fibercomplete Modules General Introduction; Principle - Viavi T-BERD/MTS 2000 User Manual

Otdr functions, options and software applications
Hide thumbs Also See for T-BERD/MTS 2000:
Table of Contents

Advertisement

Chapter 11 FiberComplete Modules
General introduction
General introduction

Principle

FiberComplete is used to perform automatically and through a single connection port,
the following tests:
Bidirectional insertion loss (IL)
Bidirectional optical return loss (ORL) using the continuous wave method
(OCWR)
Distance/length measurements
Unidirectional / Bidirectional OTDR or fault analysis
To carry out the measurements, 2x T-BERD/MTS are needed, both equipped with Fiber-
Complete capable modules (see references in
meters on the mainframes.
Using one unit at each end of the fiber under test and without any connection/disconnec-
tion, IL/ORL and distance measurements are performed and results exchanged via the
fiber under test (FUT).
If the OTDR is selected, an OTDR measurement is launched automatically:
Unidirectional mode: the OTDR acquisition is performed from the primary unit
where the test has be initiated. Traces are saved locally
Bidirectional mode: the OTDR acquisition is initiated successively from each
unit. The traces are saved locally.
A failed value of IL or ORL may trigger the Fault Finder function automatically in order
to identify the faulty event.
192
User Manual
Chapter
14) and broadband power
77000010208 Rev. 012