Manual Measurement; Typical Specifications - Viavi T-BERD/MTS 2000 User Manual

Otdr functions, options and software applications
Hide thumbs Also See for T-BERD/MTS 2000:
Table of Contents

Advertisement

Display threshold of faults:
- 0 to 5.99 dB in steps of 0.01 dB for event thresholds
- -11 to -99 dB in steps of 1 dB for the reflectance
- 0.01 to 5.99 dB in steps of 0.01 dB for attenuation
Display of slope and attenuation for a segment of fiber.
Display of the position of a fault and of attenuation.
Display of the reflectance of the fault.
Display of ORL

Manual Measurement

Measurement of slope between the cursors.
Measurement of attenuation between two segments of fiber.
Measurement of reflectance of a reflecting element.
Measurement of ORL between the two cursors.
Measurement of splice by 2 or 5 points method

Typical specifications

Typical values, measured at 25°C unless specified.
Multimode OTDR Module
Central Wavelength
Typical RMS Dynamic Range
Distance Range
Pulse width
Event Dead Zone
Attenuation Dead Zone
1. Laser in CW mode, at 25° C
2. Typical value corresponding to the difference (in dB) between the level of back-diffusion
extrapolated at the beginning of the fiber and the RMS noise level, after 3 minutes of averag-
ing, with the largest pulse width.
3. EDZ measured at +/- 1.5 dB below the peak of a non-saturated reflecting event at shortest
pulse width.
4. ADZ measured at +/- 0.5 dB on the basis of a linear regression, using a - 40 dB type reflec-
tance, at shortest pulse width.
1
2
3
4
User Manual
Chapter 13 Technical specifications
OTDR modules for T-BERD/MTS 2000/4000 V2
77000010208 Rev. 012
41XXMM
850 / 1300 nm ± 30 nm
26 / 24 dB
Up to 80 km
3 ns to 1 µs
0.8 m
4 m
233