Analysis - Viavi T-BERD/MTS 2000 User Manual

Otdr functions, options and software applications
Hide thumbs Also See for T-BERD/MTS 2000:
Table of Contents

Advertisement

Alarms > Threshold
None
User
TIA-568 C / ISO/IEC 11801 / Default
Splice Loss
Connector Loss
1
Slope
Reflectance
ORL
1. This parameter is not available in OEO-OTDR configuration
If results are above the thresholds, they will be highlighted in red in the table of results,
and the icon
If all the results lie within the thresholds (no result is in red), results are displayed in
green in the table and the icon is

Analysis

In the Setup page, press Analysis softkey (if one parameter is selected in the current
screen, press Top Menu soft key to display the right menu keys and click on Analysis).
This screen allows to configure the network:
BBU Jumper
Test from the patchcord that will be plugged into BBU.
(not applicable for DAS networks)
RRU Jumper
Test from the patchcord that will be plugged into RRU.
(not applicable for DAS networks)
Configuring the Reflectometry test for FTTA network
The alarm function is not active.
Define your own thresholds values for one or several elements:
Splice Loss / Connector Loss / Reflectance / Fiber Length Min /
Link Loss Max / ORL
Select one of this parameter to configure the alarm thresholds
with predefined values:
Default
> 0.20 dB
> 0.50 dB
> 1.00 dB/km
> - 35 dB
< 27 dB
will appear at the top right of the screen.
User Manual
Chapter 7 FTTA-SLM Software option
TIA-568C / ISO/IEC 11801 & ISO/IEC 14763-3
> 0.30 dB
> 0.75 dB
> 0.50 dB for ISO/IEC 14763-3 standard
> 1.00 dB/km
-
-
.
77000010208 Rev. 012
127