Inhalation Outlet Check Valve; Overhaul Procedures; Inlet Water Trap Filter And O-Ring Assembly; Figure 2-11 - Covidien Newport e360 Service Manual

Ventilator with accessories
Table of Contents

Advertisement

Maintenance, Overhaul, and Software Upgrade

Inhalation Outlet Check Valve

2.7.8
To replace the inhalation outlet check valve, refer to Figure 2-11.  and perform the following steps:
Remove the mixing block by removing the two retaining screws (see Figure 2-13.  ).
1.
Remove the inhalation adapter from the inhalation outlet assembly (see Figure 2-5.  ).
2.
Using your finger, push out the inhalation outlet check valve (see Figure 2-11.  ).
3.
Replace the inhalation outlet check valve diaphragm (VLV2100M).
4.
Reinstall the components and assembly by reversing steps 1-3.
5.

Overhaul Procedures

2.8
WARNING:
Disconnect electrical power, air and oxygen sources before attempting any disassembly.

Inlet Water Trap Filter and O-Ring Assembly

2.8.1
To replace the inlet filter and o-ring assemblies, follow the procedure in section 2.7.1,
Trap Filters and O-Ring
2-14
Figure 2-11.  Inhalation Outlet Check Valve
1
Inhalation outlet check valve
Assembly."
"Inlet Water
Service Manual