Output Setting - R&S NGE100B Series User Manual

Hide thumbs Also See for NGE100B Series:
Table of Contents

Advertisement

®
R&S
NGE100B
Example:

6.5.6 Output Setting

OUTPut[:STATe] <arg0>
This command defines or queries for output state for the previous selected channel.
Parameters:
<arg0>
Example:
OUTPut:GENeral <arg0>
This command activates or queries the instrument master output state.
If the master output state is activated, the Output key on the front panel will be illumi-
nated.
Parameters:
<arg0>
User Manual 5601.1343.02 ─ 04
OUTPut2 | OUTP2
Ch 2
OUTPut3 |OUTP3
Ch 3
INSTrument OUTPut1
INSTrument OUTP1
Selects the output channel as Ch 1.
INST? -> 2
Return the output channel as Ch 2.
{OFF | ON | 0 | 1}
ON | 1
Output state will be turned on.
OFF | 0
Output state will be turned off.
*RST:
OFF | 0
INST OUT1
OUTP ON
Ch 1 and output state will be activated; i.e. Ch 1 and Output key
on the front panel will be illuminated.
OUTP? -> 1
Return the output state.
{OFF | ON | 0 | 1}
ON | 1
Output state will be activated
OFF | 0
Output state will be deactivated