Firmware update via SCPI command for version 1.3 and above
1. Download the firmware file from the product homepage.
2. Unzip the firmware to the root directory of a USB flash drive.
There are two files after the unzip, do not rename these files.
3. Connect the USB flash drive to the USB port of the instrument and wait for ten sec-
onds.
4. Connect to the instrument via USB, LAN or WLAN. See
Connection",
5. Send
start firmware update remotely.
The R&S NGE100B reboots automatically as part of the update process.
6. Wait for the completion of the update process.
It takes a few minutes to complete the firmware udapte process.

5.7.7 Help

The "Help" dialog provides the internet link and QR code for the latest R&S NGE100B
information.
1. Select "Utilities > Help" menu item.
The R&S NGE100B displays the "Help" dialog. See
Figure 5-30: Help dialog
2. Use a QR code reader to scan the QR code to link to the online product home
page.
Alternatively, you could enter the address:
nge100b
User Manual 5601.1343.02 ─ 04
on page 49 and
DIAGnostic:SERVice:FWUP "NGE100FWRelease.tar.gz.enc"
into a browser to access the latest help information.