R&S NGE100B Series User Manual page 98

Hide thumbs Also See for NGE100B Series:
Table of Contents

Advertisement

®
R&S
NGE100B
TRIGger:LOGic:DIO<IO> <arg0>
This command sets or queries the trigger logic (Active High/Active Low) of the speci-
fied Digital I/O line..
Suffix:
<IO>
Parameters:
<arg0>
TRIGger:OUT:CONDition:DIO<IO> <arg0>
This command sets the trigger condition of the trigger output of the specified Digital I/O
line.
The line triggers when the output of the specified control source is turned off or on
Suffix:
<IO>
Parameters:
<arg0>
User Manual 5601.1343.02 ─ 04
CH1
Trigger input source from Ch 1.
CH2
Trigger input source from Ch 2.
CH3
Trigger input source from Ch 3.
*RST:
CH1
.
1..4
LOW | HIGH
*RST:
HIGH
.
1..4
ON | OFF | FUSetrip | CCMode | VOLTlevel | CURRlevel |
CRITevent | ARBitrary
ON
Trigger output is turned on.
OFF
Trigger output is turned off.
FUSetrip
Trigger output is linked to fuse tripped event.
VOLTlevel
Trigger output is linked to a set voltage level event.
CURRlevel
Trigger output is linked to a set current level event.
CRITevent
Trigger output is linked to a critical triggered event.
ARBitrary
Trigger output is linked to a EasyArb event.
*RST:
ON