French or Spanish spoken vitals and
general language playback is not
audible.
French or Spanish spoken vitals and
general language playback is not
working properly.
Peripheral Medical Devices
Genesis Touch - User Manual
For help with peripheral medical devices, always consult the manufac-
turer user manual that is provided with your order. One is available from
our customer portal for your convenience. Contact the manufacturer for
more information.
From the Vitals Now menu, access the Settings menu and
select About. From the About menu, confirm the Android
version number displayed. Follow the directions below,
depending on the Android version number displayed:
Android 5.1 - Follow directions in Selecting the Monitor's
Language to pick a language. With the language selected,
navigate to the Device menu, following the directions from
Enabling Device Capabilities. From the Device menu, select a
different volume (second option in the list). If a popup
appears, cancel it, and check Do Not Show Again.
Android 7.1 - Follow the directions in this manual titled,
Downloading Text-to-Speech Files and Accessing the Android
Settings menu for details. Advanced users may reference this
summary:
1. From the Android Settings menu, navigate to
Language and input.
2. Select Text-to-Speech.
3. Select the Settings button beside the Samsung
text-to-speech engine.
4. Select Install voice data.
5. Select the download icon next to the following
language first: French (Canada) - Female.
6. Accept the terms of the license agreement, and
select Start.
7. Select Install.
8. A popup may ask if you would like to regularly check 
for security problems. Select Decline to continue.
9.
When the installation finishes, select Return until
the General Settings menu appears.