Download Print this page

Select The Source State For Calibration - Agilent Technologies EMC Series Programmer's Manual

Spectrum analyzers 9 khz-1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for EMC Series:

Advertisement

Language Reference
CALibration Subsystem
Before executing this command, connect a cable between front panel connector
AMPTD REF OUT
Agilent model E7401A. If the cable is not connected, the alignment will fail.
Front Panel
Access:

Select the Source State for Calibration

:CALibration:SOURce:STATe OFF|ON|0|1
:CALibration:SOURce:STATe?
Controls the state of the 50 MHz alignment signal.
NOTE
The alignment signal is internally switched to the
E7401A. For all other models, connect a cable between front panel connector
AMPTD REF OUT
Factory Preset
and *RST:
Front Panel
Access:
Calibrate the Tracking Generator
:CALibration:TG
:CALibration:TG?
Performs an alignment of the tracking generator assembly.
The query performs the alignment and returns a zero if the alignment is successful.
This command is applicable on all Agilent EMC models except E7401A with
NOTE
Option 1DN installed. Before executing this command, connect a cable between
front panel connector
using command CAL:TG if the cable is not connected.
Front Panel
Access:
232
and the
connector for all Agilent EMC analyzers except
INPUT
System, Alignments, Align Now, RF
and the
connector before performing a calibration.
INPUT
Off
For Agilent EMC model E7401A:
Input/Output (or Input), Amptd Ref (f=50 MHz) On Off
For all other Agilent EMC models:
Input/Output (or Input), Amptd Ref Out (f=50 MHz) On Off
and the
RF OUT
INPUT
System, Alignments, Align Now, TG
INPUT
for Agilent model
connector. The alignment will fail
Chapter 5

Advertisement

loading

This manual is also suitable for:

E7404aE7402aE7405aE7401aE7403a