Download Print this page

Ndbstate - Agilent Technologies EMC Series Programmer's Manual

Spectrum analyzers 9 khz-1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for EMC Series:

Advertisement

Language Reference
CALCulate Subsystem

NdBstate

:CALCulate:BWIDth|BANDwidth[:STATe] OFF|ON|0|1
:CALCulate:BWIDth|BANDwidth[:STATe]?
Controls the bandwidth measurement function. The function measures the
bandwidth, at the number of dB down specified in :CALCulate:BWIDth:NDB,
of the maximum signal on the display.
Factory Preset
and *RST:
Off
Remarks:
When this command is turned on, the bandwidth measurement
function (N dB Points) is associated with the active marker. If
no marker is active at the time this command is turned on,
marker 1 becomes the active marker, and a peak search is
performed. No restrictions exist for moving the bandwidth
measurement function markers to any other signal on the
display. However, when this function is turned on, all other
concurrent marker functions are suspended.
Front Panel
Access:
Peak Search (or Search), N dB Points On Off
Calculate Correction at Frequency
:CALCulate:CORRection:ATFREquency? <freq>
Calculates the total correction factor for a specified frequency point.
Example:
:CALC:CORR:ATFRE? 1.5E8
Remarks:
Refer to [:SENSe]:CORRection:CSET[1]|2|3|4:DATA
<freq>, <rel_ampl> {,<freq>, <rel_ampl} for an
explanation of correction factors.
Test Current Trace Data Against all Limit Lines
:CALCulate:CLIMits:FAIL?
Queries the status of the limit line testing. Returns a 0 if the trace data passes when
compared with all the current limit lines. Returns a 1 if the trace data fails any limit
line test.
200
Chapter 5

Advertisement

loading

This manual is also suitable for:

E7404aE7402aE7405aE7401aE7403a