Language Reference
CALCulate:LLINe Subsection

Test the Data Against the Limit Line

:CALCulate:LLINe[1]|2:FAIL?
Queries the status of the limit line testing. Returns a 0 if the data passes, and
returns a 1 if there is a failure. This query value is valid only if margin or limit test
is On. Use the command :CALCulate:LLINe[1]|2:STATe OFF|ON|0|1 to
activate limit line testing.
Set the Margin Size
:CALCulate:LLINe[1]|2:MARGin <rel_ampl>
:CALCulate:LLINe[1]|2:MARGin?
Allows you to define the amount of measurement margin that is added to the
designated limit line.
Factory Preset
and *RST:
not affected
Default Units:
dB
Remarks:
The margin must be negative for upper limit lines, and positive
for lower limits.
Front Panel
Access:
Display, Limits, Limit 1|2, Margin On Off
Display the Limit Margin
:CALCulate:LLINe[1]|2:MARGin:STATe OFF|ON|0|1
:CALCulate:LLINe[1]|2:MARGin:STATe?
Allows you to display a measurement margin that is added to the designated limit
line to do secondary testing of the data.
Factory Preset
and *RST:
Off
Front Panel
Access:
Display, Limits, Limit 1|2, Margin On Off
214