Coarse Adjust the Frequency Reference

:CALibration:FREQuency:REFerence:COARse <setting>
:CALibration:FREQuency:REFerence:COARse?
Allows coarse adjustment of the internal 10 MHz reference oscillator timebase of
the analyzer.
:CALibration:ALL is required after COARse is set.
NOTE
Range:
Front Panel
Access:
Fine Adjust the Frequency Reference
:CALibration:FREQuency:REFerence:FINE <setting>
:CALibration:FREQuency:REFerence:FINE?
Allows fine adjustment of the analyzer internal 10 MHz reference oscillator
timebase.
NOTE
:CALibration:ALL is required after FINE is set.
Range:
Front Panel
Access:
Select the Frequency Corrections
:CALibration:FREQuency[:STATe] OFF|ON|0|1
:CALibration:FREQuency[:STATe]?
Turns on or off the frequency corrections.
Factory Preset
and *RST:
Front Panel
Access:
Align the RF Circuitry
:CALibration:RF
:CALibration:RF?
Performs an alignment of the RF assembly.
The query performs the alignment and returns a zero if the alignment is successful.
Chapter 5
Integer, 0 to 255
System, Alignments, Time Base, Coarse
Integer, 0 to 255
System, Alignments, Time Base, Fine
On
System, Alignments, Freq Correct On Off
Language Reference
CALibration Subsystem