Use Status Registers to Determine the State of Analyzer Events and Conditions
Bit
Decimal
Value
7
128
8
256
9
512
10
1024
11
2048
12
4096
13
8192
14
16384
15
32768

Questionable Status Power Condition and Event Registers

The Questionable Status Power Condition Register continuously monitors output
power status of the analyzer. Condition registers are read-only. To query the
condition register, send the command
:STATus:QUEStionable:POWer:CONDition? The response will be the
decimal sum of the bits which are set to 1.
The transition filter specifies which types of bit state changes in the condition
register will set corresponding bits in the event register. The changes may be
positive (from 0 to 1) or negative (from 1 to 0). Send the command
:STATus:QUEStionable:POWer:NTRansition <num> (negative
transition) or :STATus:QUEStionable:POWer:PTRansition <num>
(positive transition) where <num> is the sum of the decimal values of the bits you
want to enable.
The Questionable Status Power Event Register latches transition events from the
condition register as specified by the transition filters. Event registers are
destructive read-only. Reading data from an event register will clear the content of
that register. To query the event register, send the command
:STATus:QUEStionable:POWer[:EVENt]?
See
"Questionable Status Event Enable Register" on page 77
how to set the summary bit using the event enable register. In this case, use the
command :STATus:QUEStionable:POWer:ENABle <num>.
Chapter 2
Description
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
CALibration Summary: This is the summary bit for the
Questionable Calibration Status Register.
INTegrity Sum: This is the summary bit for the
Questionable Integrity Status Register.
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
Always Zero (0)
Status Registers
for an explanation of