Language Reference
[:SENSe]:CORRection Subsection

Delete Amplitude Correction

[:SENSe]:CORRection:CSET[1]|2|3|4:DELete
Deletes the specified correction set. If the set was On, it is turned Off.
Front Panel
Access:
Set Amplitude Correction Frequency Interpolation
[:SENSe]:CORRection:CSET[1]|2|3|4:X:SPACing
LINear|LOGarithmic
Sets the frequency interpolation to linear or logarithmic for the specified correction
set.
Remarks:
Front Panel
Access:
Perform Amplitude Correction
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]?
Turns the amplitude correction function on or off for the given set.
NOTE
[:SENSe]:CORRection:CSET:ALL[:STATe] must be on for this command to
function.
Factory Preset
and *RST:
Remarks:
Front Panel
Access:
278
AMPLITUDE/Y Scale, Corrections,
Antenna|Cable|Other|User, Delete Correction
Logarithmic frequency scale corrections are linearly
interpolated between correction points with respect to the
logarithm of the frequency. Linear frequency scale corrections
are interpolated along straight lines, connecting adjacent points
on a linear scale.
AMPLITUDE/Y Scale, Corrections, Freq Interp Log Lin
Off
CSET number equivalents to front panel access definitions are
as follows:
CSET or CSET1 is Antenna
CSET2 is Cable
CSET3 is Other
CSET4 is User
AMPLITUDE/Y Scale, Corrections,
Antenna|Cable|Other|User, Correction On Off
Chapter 5