Time Gate Level (Option 1D6 Only) - Agilent Technologies EMC Series Programmer's Manual

Spectrum analyzers 9 khz-1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for EMC Series:

Advertisement

Time Gate Level (Option 1D6 Only)

[:SENSe]:SWEep:TIME:GATE:LEVel HIGH|LOW
[:SENSe]:SWEep:TIME:GATE:LEVel?
Selects the level of the gate signal; this command is for LEVel triggering only.
Factory Preset
and *RST:
Front Panel
Access:
Time Gate Polarity (Option 1D6 Only)
[:SENSe]:SWEep:TIME:GATE:POLarity NEGative|POSitive
[:SENSe]:SWEep:TIME:GATE:POLarity?
Selects the polarity of the gate signal; this command is for EDGE triggering only.
Factory Preset
and *RST:
Front Panel
Access:
Preset Time Gate (Option 1D6 Only)
[:SENSe]:SWEep:TIME:GATE:PRESet
Presets the time-gated spectrum analysis capability.
Remarks:
Control Time Gate (Option 1D6 Only)
[:SENSe]:SWEep:TIME:GATE[:STATe] OFF|ON|0|1
[:SENSe]:SWEep:TIME:GATE[:STATe]?
Turns time gating on or off.
Time gate cannot be turned on if external trigger delay is on.
NOTE
Chapter 5
High
Sweep, Gate Setup, Level Setup
Positive
Sweep, Gate, Edge Gate, Slope Pos Neg
This command resets gate parameters to default values, as
follows:
Gate trigger type = edge
Gate polarity = positive
Gate delay = 1 µs
Gate length = 1 µs
Gate level = high
Language Reference