Preset - Agilent Technologies EMC Series Programmer's Manual

Spectrum analyzers 9 khz-1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for EMC Series:

Advertisement

Language Reference
SYSTem Subsystem
History:
Added with firmware revision A.04.00.
Remarks:
Disable the output ports for faster measurement times.

Preset

:SYSTem:PRESet
Returns the instrument to a set of defined conditions. The particular set is selected
by :SYSTem:PRESet:TYPE. This command does not change any persistent
parameters. The term persistent means that the command retains the setting
previously selected, even through a power cycle.
Front Panel
Access:
Preset
Persistent State Reset
:SYSTem:PRESet:PERSistent
Sets the persistent state values to their factory defaults. The term persistent means
that the command retains the setting previously selected, even through a power
cycle. Examples of persistent functions are: GPIB address, power-on type, and
preset type.
Front Panel
Access:
System, Restore Sys Defaults
Preset Type
:SYSTem:PRESet:TYPE FACTory|USER|MODE
Selects the preset state to be either factory-defined or user-defined preset
conditions.
Factory Preset
and *RST:
The factory default is MODE. This parameter is persistent, which
means that it retains the setting previously selected, even
through a power cycle.
History:
Changed with firmware revision A.08.00. Previous firmware
revisions had default type FACTory.
Remarks:
:SYSTem:PRESet:USER:SAVE defines the user preset.
Front Panel
Access:
System, Power On/Preset, Preset Type