[:SENSe]:AVERage Subsection

Clear the Current Average
[:SENSe]:AVERage:CLEar
Re-start the trace averaging function.
Re-start the trace at the beginning of a sweep to obtain valid average data. To
NOTE
do this, remotely abort the sweep and initiate a single sweep.
Set the Average Count
[:SENSe]:AVERage:COUNt <integer>
[:SENSe]:AVERage:COUNt?
Specifies the number of measurements that are combined.
Factory Preset
and *RST:
Range:
Front Panel
Access:
Turn Averaging On/Off
[:SENSe]:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:AVERage[:STATe]?
This command toggles averaging off and on. Averaging combines the value of
successive measurements to average out measurement variations.
Factory Preset
and *RST:
Remarks:
Front Panel
Access:
Chapter 5
100
1 to 8192
BW/Avg, Average On Off
Off
When a measurement under the front panel
started, this command is turned off for video averaging
([:SENSe]:AVERage:TYPE VIDeo). If this command is
turned on for video averaging when any of the
measurements are in progress, that measurement will be
stopped.
BW/Avg, Average On Off
Language Reference
[:SENSe]:AVERage Subsection
key is
MEASURE
key