Sweep Time - Agilent Technologies EMC Series Programmer's Manual

Spectrum analyzers 9 khz-1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for EMC Series:

Advertisement

Factory Preset
and *RST:
Linear
Remarks:
Refer to the User's Guide for detailed information on the
interactions of this command with other functions.
History:
Added with firmware revision A.08.00.
Front Panel
Access:
FREQUENCY, Scale Type

Sweep Time

[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
Specifies the time in which the instrument sweeps the display.
Factory Preset
and *RST:
120.8 ms
Range:
The range depends upon the installed options, number of sweep
points, and firmware revision of your instrument. See "Sweep
Time Range" in the Specifications Guide for details.
Default Unit:
seconds
Remarks:
A span value of 0 Hz causes the analyzer to enter zero span
mode. In zero span the X-axis represents time rather than
frequency. In this mode, the sweep time may be set to faster
values when Option AYX is installed.
Front Panel
Access:
Sweep, Sweep Time Auto Man
Automatic Sweep Time
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TIME:AUTO?
Automatically selects the fastest sweep time for the current settings.
Factory Preset
and *RST:
On
History:
This command function changed with firmware revision
A.08.00. With :AUTO ON in zero span, an error will be
generated.
Front Panel
Access:
Sweep, Sweep Time Auto Man
Chapter 5
Language Reference
[:SENSe]:SWEep Subsection
299