Download Print this page

Define Peak Threshold - Agilent Technologies EMC Series Programmer's Manual

Spectrum analyzers 9 khz-1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for EMC Series:

Advertisement

Access:

Define Peak Threshold

:CALCulate:MARKer:PEAK:THReshold <ampl>
:CALCulate:MARKer:PEAK:THReshold?
Specifies the minimum signal level for the analyzers internal peak identification
routine to recognize a signal as a peak. This applies to all traces and all windows.
See command :CALCulate:MARKer:PEAK:SEARch:MODE
NOTE
Range:
Default Unit:
Front Panel
Access:
Peak to Peak Delta Markers
:CALCulate:MARKer[1]|2|3|4:PTPeak
Positions delta markers on the highest and lowest points on the trace.
Factory Preset
and *RST:
Front Panel
Access:
Set Center Frequency to the Marker Value
:CALCulate:MARKer[1]|2|3|4[:SET]:CENTer
Sets the center frequency equal to the specified marker frequency, which moves the
marker to the center of the screen. In delta marker mode, the center frequency is set
to the marker delta value. This command is not available in zero span.
Front Panel
Access:
Set Reference Level to the Marker Value
:CALCulate:MARKer[1]|2|3|4[:SET]:RLEVel
Sets the reference level to the specified marker amplitude. In delta marker mode,
the reference level is set to the amplitude difference between the markers.
Front Panel
Access:.
.
Chapter 5
Peak Search (or Search), Search Criteria, Peak Search
Type, Max Value|Excursion & Threshold
Reference level to the bottom of the display
Amplitude units
Peak Search (or Search), Search Criteria, Peak Threshold
Off
Peak Search (or Search), Pk-Pk Search
Marker –>, Mkr –> CF
Marker –>, Mkr –> Ref Lvl
Peak Search (or Search), Meas Tools, Mkr –> Ref Lvl
Language Reference
CALCulate:MARKer Subsection