Questionable Integrity Negative Transition

:STATus:QUEStionable:INTegrity:NTRansition <integer>
:STATus:QUEStionable:INTegrity:NTRansition?
This command determines which bits in the Questionable Integrity Condition
register will set the corresponding bit in the Questionable Integrity Event register
when that bit has a negative transition (1 to 0). The variable <integer> is the sum of
the decimal values of the bits that you want to enable.
Factory Preset
and *RST:
0
Range:
Integer, 0 to 32767
Questionable Integrity Positive Transition
:STATus:QUEStionable:INTegrity:PTRansition <integer>
:STATus:QUEStionable:INTegrity:PTRansition?
This command determines which bits in the Questionable Integrity Condition
register will set the corresponding bit in the Questionable Integrity Event register
when that bit has a positive transition (0 to 1). The variable <integer> is the sum of
the decimal values of the bits that you want to enable.
Factory Preset
and *RST:
32767 (all 1's)
Range:
Integer, 0 to 32767
Questionable Integrity Uncalibrated Enable
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle
<integer>
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle?
This command determines which bits in the Questionable Integrity Uncalibrated
Condition Register will set bits in the Questionable Integrity Uncalibrated Event
register, which also sets the Data Uncalibrated Summary bit (bit 3) in the
Questionable Integrity Register. The variable <integer> is the sum of the decimal
values of the bits you want to enable.
Factory Preset
and *RST:
32767 (all 1's)
Range:
Integer, 0 to 32767